Sign In | Join Free | My chinacsw.com |
|
Brand Name : KOMEG
Model Number : KTS-72B
Certification : CE, ISO,
Place of Origin : China
MOQ : 1SET
Price : negotiation
Payment Terms : T/T, negotiable
Supply Ability : 1000 SETS / YEAR
Delivery Time : 45 working days after received your payment
Packaging Details : Packed in carton firstly, and then reinforced with plywood case for outer packing
Chamber Model : KTS Series
Chamber Volume Size : 27L to 480L
Shock range : -55ºC~+150ºC
Maximal Temperature : 200°C
Lowest Temperature : -70°C
Transition time : ≤15S
Recovery time : 5min
Temperature change rate : 15 to 50°C/min
High Stability Explosion-proof Lock 72L Three-Zone Type Temperature Impact Chamber
1. Description and Application
As we know that thermal shock occurs when a thermal gradient causes different parts of an object to expand by different amounts. This differential expansion can be understood in terms of stress or of strain, equivalently. At some point, this stress can exceed the strength of the material, causing a crack to form. If nothing stops this crack from propagating through the material, it will cause the object's structure to fail. Normally we use thermal shock test equipment to test how much amount can a product withstand thermal shock.
Thermal shock testing exposes products to alternating low and high temperatures to accelerate failures caused by temperature cycles or thermal shocks during normal use. The transition between temperature extremes occurs very rapidly, greater than 15 °C/min.
2. Working Principle
There are three chambers used to perform thermal shock testing, high-temperature chamber, low-temperature chamber and test chamber. Sample is placed in the test chamber, more extreme temperature than test temperature can be set in high-temperature chamber and low-temperature chamber. When doing low temperature testing, cold chamber door open, and low-temperature chamber working together with test chamber.
When converted into high-temperature test, cold chamber door closed, hot chamber door are opened, and test chamber working together with high-temperature chamber. Conversion of mechanical action (transferring from high temperature to low temperature or low temperature to high temperature) can be completed in less than 1second, and the temperature can be quickly stabilized.
During the whole test, test sample is no need to be moved, and without any human intervention.
3. Specification
(1) KTS-A Series
Model |
| ||||||||
72A | 100A | 150A | 200A | 252A | 300A | 480A | |||
Inside size(mm) | W | 450 | 500 | 600 | 670 | 700 | 800 | 800 | |
H | 400 | 450 | 500 | 600 | 600 | 600 | 800 | ||
D | 400 | 450 | 500 | 500 | 600 | 650 | 750 | ||
Outside size(mm) | W | 1490 | 1540 | 1640 | 1720 | 1750 | 1850 | 1850 | |
H | 1790 | 1840 | 1890 | 2000 | 2000 | 2000 | 2200 | ||
D | 1600 | 1700 | 1830 | 1850 | 1930 | 1980 | 2500 | ||
Test weight | 5 | 10 | 12 | 15 | 18 | 20 | 20 | ||
Low Temp Chamber | -55ºC~-10ºC, Pre cooling time: ≤ 60 min from +20ºCto-55ºC | ||||||||
Test Chamber | -40ºC~+150ºC | ||||||||
High Temp Chamber | +60ºC~+200ºC ,Within 20 min, Pre-heating time | ||||||||
Transition time | ≤15S | ||||||||
Recovery time | High-Temp Exposure 30 min to reach & stabilize required hot temp of test sample | ||||||||
Low-Temp Exposure 30 min to reach & stabilize required cold temp of test sample | |||||||||
≤ 5 min | |||||||||
Ambient Temp. | +5ºC~+35ºC | ||||||||
Power | AC 380±10% 50Hz ,3 phase 4 wires +Ground Wires |
(2) KTS-B Series
Model |
| ||||||||
72B | 100B | 150B | 200B | 252B | 300B | 480B | |||
Inside size(mm) | W | 450 | 500 | 600 | 670 | 700 | 800 | 800 | |
H | 400 | 450 | 500 | 600 | 600 | 600 | 800 | ||
D | 400 | 450 | 500 | 500 | 600 | 650 | 750 | ||
Outside size(mm) | W | 1490 | 1540 | 1640 | 1720 | 1750 | 1850 | 1850 | |
H | 1790 | 1840 | 1890 | 2000 | 2000 | 2000 | 2200 | ||
D | 1600 | 1700 | 1830 | 1850 | 1930 | 1980 | 2500 | ||
Test weight(KG) | 5 | 10 | 12 | 15 | 18 | 20 | 20 | ||
Low Temp. chamber | -70ºC~-10ºC ,Pre cooling time: +20ºC~-70ºC ,≤ 60 min | ||||||||
Test. Chamber | -55ºC~+150ºC | ||||||||
High Temp. chamber temperature | +60ºC~+200ºC,Preheating time: ≤20 min | ||||||||
Transition time | ≤15S | ||||||||
Recovery time | High-Temp Exposure 30 min to reach & stabilize required hot temp of test sample | ||||||||
Low-Temp Exposure 30 min to reach & stabilize required cold temp of test sample | |||||||||
≤ 5 min | |||||||||
Ambient Temp. | +5ºC~+35ºC | ||||||||
Power | AC 380±10% 50Hz ,3 phase 4 wires +Ground Wires |
(3) KTS-D Series
Model |
| ||||||||
72D | 100D | 150D | 200D | 252D | 300D | 480D | |||
Inside size(mm) | W | 450 | 500 | 600 | 670 | 700 | 800 | 800 | |
H | 400 | 450 | 500 | 600 | 600 | 600 | 800 | ||
D | 400 | 450 | 500 | 500 | 600 | 650 | 750 | ||
Outside size(mm) | W | 1490 | 1540 | 1640 | 1720 | 1750 | 1850 | 1850 | |
H | 1790 | 1840 | 1890 | 2000 | 2000 | 2000 | 2200 | ||
D | 1600 | 1700 | 1830 | 1850 | 1930 | 1980 | 2500 | ||
Test weight | 5 | 10 | 12 | 15 | 18 | 20 | 20 | ||
Low Temp Chamber | -80ºC~-10ºC, Pre cooling time: ≤ 60 min from +20ºCto-80ºC | ||||||||
Test Chamber | -65ºC~+150ºC | ||||||||
High Temp Chamber | +60ºC~+200ºC ,Within 20 min, Pre-heating time | ||||||||
Transition time | ≤15S | ||||||||
Recovery time | High-Temp Exposure 30 min to reach & stabilize required hot temp of test sample | ||||||||
Low-Temp Exposure 30 min to reach & stabilize required cold temp of test sample | |||||||||
≤ 5 min | |||||||||
Ambient Temp. | +5ºC~+35ºC | ||||||||
Power | AC 380±10% 50Hz ,3 phase 4 wires +Ground Wires |
4. Product Picture:
![]() |
High Stability Explosion proof 72L Temperature Impact test equipment Images |